|
AAS |
Principle of Atomic Absorption Spectrometer
- A solution of metallic
salt is sprayed on a flame, and fine droplets are formed.
- The thermal energy of flame and the solvents in the droplets evaporate, leaving a fine residue, which is
converted to neutral atoms.
- The neutral atoms
absorb radiation of a specific wavelength emitted by a hollow cathode
lamp(HCL).
- HCL is filled with the
vapour of elements which gives a specific wavelength of radiation.
- The intensity of light
absorbed by neutral atoms is directly proportional to the concentration of the
elements.
- The intensity of
radiation absorbed by neutral atoms is measured using photometric detectors.
- Note: The excitation of neutral atoms is
brought only by radiation from a hollow cathode lamp and not by the thermal
energy of the flame.
Instrumentation of Atomic Absorption Spectrometer
Hollow Cathode Lamp
- The source of light in
AAS is a hollow cathode lamp.
- The cathode is made up
of specific elements or alloys of elements on the cathode.
- When the current is
applied to the anode and cathode, metal ions emerge from the collide with filler
gas, which is argon. Due to these
collisions, the number of metal atoms is excited and emit their characteristic
radiation.
- This radiation is
absorbed by neutral atoms of the same elements in the ground state, which occurs
in the flame when the sample solution is sprayed.
- For the determination of
every element, separate HCL has to be using multi-element lamps.
- Ex: Two elements lamps
Na/K, Ca/Mg, Three elements lamps Ca/Mg/ Zn.
Burner
- There are different
burners available, which are used to spray the sample solution into fine droplets and mix it with fuel and oxidant.
- The most common ones are the Total
consumption burner and the Laminar flow burner.
Chopper
- The instrument had
choppers, which rotated as fans allowing alternatively radiation from flame alone
or radiation from HCL and flame.
- This procedure is a
pulsating current signal, which is used to measure the intensity of light
absorbed by elements, without interference by radiation from the flame itself.
Monochromator
- To isolate the line
spectrum of the element from the background signal of the flame. So
monochromator that can provide a good resolution of 1 nm is required.
Detector
- The intensity of
radiation absorbed by elements, in the UV or Visible region (190-800
nm) can be detected phototube or a photomultiplier tube.
Recorder
- The recorder readout
is capable of displaying the absorption spectrum as well as absorbance at a
specified wavelength.
Application of Atomic Absorption Spectrometer
- Atomic Absorption
Spectrometer is mainly used for quantitative analysis of various elements
present in different samples.
- Major of the
applications are Agriculture, Forensic, Environmental, Food, Pharma,
Mining, and petrochemical industries.
- Estimation of trace
elements in biological fluids (ex: blood, urine, etc).
- Estimation of elements
like copper, Nickel and Zinc, Magnesium, Zinc, Mercury, Lead, etc.
- Estimation of elements
in soil samples, water supply, effluent ceramics, etc.
SOP for Atomic Absorption Spectrometer
- Company
Logo Company Name, Unit, Address, Area
TITLE: Operation and
Calibration of Atomic Absorption Spectrometer
SOP No. QC-XX, Department: Quality Control Revision No.0.0
. Page No.xx of Effective Date: xx. yy. Z Review Date: 3 yrs from the effective date
PURPOSE
- This procedure
describes the Operation and Calibration of the Atomic Absorption Spectrometer
SCOPE
- This procedure applies to the QC departments at the Company name, Unit, and Address.
ABBREVIATIONS
- AAS: Atomic Absorption Spectrophotometer.
- FAAS: Flame Atomic
Absorption Spectrophotometer.
- GFAAS: Graphite
Furnace Atomic Absorption Spectrophotometer.
- VGA: Vapour Generation
Accessory.
- Abs: Absorbance
- nm: Nanometer
- ppb: Parts per billion
- ppm: Parts per million
- RSD: Relative Standard
Deviation
- NLT: Not Less Than.
- NMT: Not More Than
- QC: Quality control
- QA: Quality Assurance
RESPONSIBILITY
- Analysts/Users are
responsible for following this procedure.
- QC Manager / Designee
is responsible for the implementation of this procedure.
- QA Manager / Designee
shall monitor compliance with this procedure.
Safety Precautions
- Operate the
spectrophotometer with the flame shield closed and the sample compartment front
panel in place.
- Use the correct burner
head for different flame modes.
- Do not ignite the flame
without a burner head and nebulizer and also when the water level is too low in
the liquid trap.
- Do not use flammable
solvents in or near the AAS which can create an explosion.
- Always use the correct
acetylene flow for Nitrous oxide – Acetylene mode.
- This mode will produce
carbon deposits, making sure the extraction equipment above the instrument is
working correctly.
- Always empty and clean
the liquid trap after completion of analysis.
- Never use a solvent
having a weight per ml lower than 0.76.
Operation of Atomic Absorption Spectrometer
Flame Mode
- Switch on the
instrument. Check all the tubings for gas leaks.
- Select the gas either air-acetylene or nitrous oxide-acetylene as per the method.
- Place the hollow
cathode lamp, mention the current, and perform an optical set-up.
- Ignite the flame and
allow it for a few minutes to stabilize.
- Carry out adjustments
for flame height, gas flow rate, and burner position to obtain maximum
absorption signal. Ensure that a real-time graph is displayed.
- Aspirate deionized
water and auto-zero the absorbance signal.
- Click the start icon
followed by calibration and sequence.
GFAAS Mode
- Switch on the
instrument. Check all gas leaks.
- Turn on the hollow
cathode lamp, and/or perform an optical setup, and allow the lamp to warm up. Switch on the cooling
water supply and argon gas with an exit pressure of 70 psi.
- Set furnace heat
program.
- Choose the composition
of the wash liquid to efficiently remove traces of the sample solutions from
the capillary.
- Dilute nitric acid (approximately 1% v/v) will be suitable
for many aqueous sample types.
- The autosampler capillary
tip must be aligned with the cuvette injection hole before use.
- Set the Argon gas
supply. Turn on the water supply.
VGA Mode
- Switch on the
spectrophotometer install the Hydride generator parts and connect.
- Turn on the hollow
cathode lamp, perform an optical setup, and allow the lamp to warm up.
- Prepare the reagents,
sample as required, and place them in the containers
- Set the heating
element and quartz tube on the burner head.
- Set the temperature
for the hydride generator. Connect the Argon tubing.
- Define sequence and start
analyzing.
Startup AAS Software
⇓
Turn on power
⇓
Run mode
⇓
Analysis
⇓
Initialization
⇓
Set the Element lamp, Select the element a, and Set the Element
Lamp
.⇓
Peak search and wavelength scan
⇓
Set the instrumental parameters ( Optical and measure method )
⇓
Set the elemental measurement parameters
.⇓
Optimize the atomizer position for furnace mode
.⇓
Set the sample
⇓
Standard concentration set-up
⇓
Select the Correction Method, Curve equation, concentration
unit, sample
.⇓
Select the measurement parameters and perform the measurement
.
Calibration of Atomic Absorption Spectrometer
Flame
System Sensitivity Check
- Measure 2.0 ppm Copper standard solution.
Acceptance Criteria:
- The average of seven
absorbance readings should be NLT 0.280 Abs&% RSD should be NMT 1.
Linearity in Flame Mode
- Perform Linearity with at least three calibration
points for Zinc (213.9 nm), Copper (324.7 nm), and K (766.5 nm).
Acceptance Criteria: Regression should be NLT 0.99
GFAAS Sensitivity Check
- Measure 20 µL
injection of 20 ppb Copper standard solution
Acceptance Criteria:
- The average of seven
absorbance readings should be NLT 0.06 Abs & % RSD should be NMT 5.
- Perform Linearity with
at least three calibration points for Lead (283.3 nm)
Acceptance Criteria:
Regression should be NLT 0.99
VGA MODE (Hydride Generator)
- Linearity: Perform
Linearity with at least three calibration points for Mercury.
Acceptance Criteria:
Regression should be NLT 0.99.
Linearity in Flame Mode
- Perform Linearity with
at least three calibration points for Copper (324.7 nm)Acceptance Criteria:
Regression should be NLT 0.99
Linearity in GFAAS ModeLinearity:
- Perform
Linearity with at least three calibration points for Mercury. Acceptance
Criteria: Regression should be NLT 0.99Perform Linearity with at least three
calibration points for Lead (283.3 nm)Acceptance Criteria: Regression should be
NLT 0.99
VGA MODE (Hydride
Generator)
- Calibration
Frequency: Once in six months.
Maintenance:
- Check the exhaust
system using the smoke.
- Check the acetylene
cylinder pressure.
- Ensure that the
acetylene is not drawn into the instrument.
- Inspect the gas hoses
for any damage.
- Turn on the gas
supplies to the instrument and check all hoses and connections for leaks.
- Check water
connections for GFAAS operation. Check the burner for deposits.
- If any, soak
with detergent and clean with distilled water.
- At frequent intervals
inspect the inside of the graphite shroud which is mounted at the center cooling block.
- Using a cotton bud or
similar material, ensure that the bore and injection hole are free from loose
carbon or sample residues.
REFERENCES
REVISION HISTORY
- Version no. Effective
date Summary of revision
0.0 xx. yy.zzzz New SOP
END OF THE DOCUMENT
Prepared by Reviewed
by Approved by
Name:
Designation: Executive-QC Manager-QC Manager-QA
Signature Date: