
Auto Polarimeter
STANDARD
OPERATING PROCEDURE
Company Logo |
Name of the Company Unit, Address, City |
||
TITLE
|
Operation and Calibration of Auto Polarimeter |
||
SOP
No. |
QC-0XX |
Department |
Quality
Control |
Revision
No. |
0.0 |
Page No. |
1 of 10 |
Effective
Date |
XX/YY/ZZZZ |
Review Date |
XX/YY/ZZZZ |
1.0 PURPOSE
1.1 This procedure describes the Operation and Calibration of an Auto Polarimeter.
2.0 SCOPE
2.1
This procedure applies to Pharma
Lab at Name of the company, Address.
3.0 ABBREVIATIONS/ DEFINITIONS
3.1
USP - United States Pharmacopoeia
3.2
MCP- Modular circular polarimeter
3.3 Optical Rotation:
The optical rotation of a substance is the angle through which the plane of polarization is rotated when polarized passes through the substance, if liquid, or a solution of the substance, if solid Substances are described as dextrorotatory or laevorotatory according to whether the plane of polarization is rotated clockwise or anti-clockwise respectively as determined by viewing the light source. Dextrorotatory is designated (+) and laevorotatory is designated (-). Optical rotation is expressed in degrees.
3.4 Specific Optical Rotation:
The specific optical rotation [α]D25 of a substance is the angle of rotation, ‘α’, of the plane of polarization at the wavelength of the D line of sodium (λ=589.3) measured at 250C unless otherwise specified, calculated regarding a 1dm thick layer of a solution containing 1 gm of the substance per mL. It is always expressed regarding a given solvent and concentration.
4.0
RESPONSIBILITY
4.1 The user is responsible for following this
procedure.
4.2 QC HOD / Designee shall be responsible for the implementation of this procedure.
4.3 QA
shall monitor for compliance with this procedure.
5.0 PROCEDURE
5.1 Operation
5.1.1 Ensure
that the working bench and instrument are clean.
5.1.2 Switch
“ON” the instrument
5.1.3 Log in: Click on the user indicator (right upper corner), and the login dialogue box
opens.
5.1.4 Enter the user name and password on the login screen
5.1.5 Enter by pressing <OK>
5.1.6 After logging in, the main screen appears with the
following operational functions.
·
Header:
Contains current method name and sample
·
Content
area: This shows the measuring value.
·
Output
field: Shows output parameters.
·
Progress
bar: Shows the actual status of the measurement
·
Toolbar: Consists of the menu, measured data, Quick setting, method and start
5.1.7 On the right side of the main screen there is a quick
access area that displays the general instrument status as well as all measuring
errors that have occurred during the measurement. The button colour changes its appearance to
green, yellow and red depending on the current error status.
5.1.8 Green sign Indicates: The general instrument status is OK all measurements are ok
5.1.10 Red flash sign Indicates:
The instrument has a major problem.
One or more samples could not be measured.
5.1.11
To
reset the diagnosis button to the green sign, confirm all error messages by
tapping on the <x> button on the right side of the message.
5.1.12
Wait
for min.15 minutes, to initialise the instrument.
5.1.13
Thoroughly
flush the sample cell and fill it bubble-free with an optically inactive liquid
(usually the sample solvent).
5.1.14
Position
the measurement cell in the measuring compartment and close the cover.
5.1.15
Tap
the optical button on the right side of the main screen.
5.1.16
The
optical zero point is set and the display shows zero value for the optical
rotation.
5.1.17
Remove
the solvent from the sample cell.
5.1.18
Fill
in the Sample homogenously bubble-free & avoid solid particles because they
would affect the measurement result.
5.1.19
Click
on start. Enter the fields such as sample name; batch No. Concentration and
temperature again click on start.
5.1.20
The
instrument starts the measurement. When the measurement is over the message
“finished “appears on the screen.
5.1.21
Repeat
the steps 5.1.15 to 5.1.19 for the consecutive measurement of the samples
5.2
Calibration
5.2.1
Temperature calibration:
· Check
the sensor at 20°C and 25°C using a certified calibrated thermometer
and allow sufficient time for thermal equilibrium.
· Fill
the sample cell with water and set the desired temperature. Insert the probe of the
thermometer in the sample cell
· Compare
the display temperature with a certified thermometer under equilibrium conditions.
The difference should be Not more than ±0.5°C.
5.2.2 Accuracy and Linearity with Sucrose
Solution:
· Optical
rotation of the solutions should be determined within 30 minutes of
preparation.
· All
the measurements in the calibration activity should be taken at 25± 0.50C
at 589nm.
· Use the same cell for the sample and test solution.
· Place the cell that the light passes through in the same direction each time.
Preparation of 10% w/v Sucrose solution:
·
Weigh about 10 g of sucrose and transfer
it into a 100 mL volumetric flask.
·
Dissolve and make to the volume with
water.
Preparation of 20% w/v Sucrose
solution:
·
Weigh about 20 g of sucrose and transfer
it into a 100 mL volumetric flask.
·
Dissolve and make to the volume with
water.
Preparation of 30% w/v Sucrose
solution:
·
Weigh about 30 g of sucrose and transfer
it into a 100 mL volumetric flask.
·
Dissolve and make to the volume with
water.
Preparation of 40% w/v Sucrose
solution:
·
Weigh about 40 g of sucrose and transfer
it into a 100 mL volumetric flask.
·
Dissolve and make to the volume with
water.
Preparation of 50% w/v Sucrose
solution:
·
Weigh about 50 g of sucrose and transfer
it into a 100 mL volumetric flask.
·
Dissolve and make to the volume with
water.
5.2.3
Procedure
·
Allow the instrument to warm up for 30
minutes.
·
Take 100mm Polarimeter cell and rinse
and fill with water which is at 25 ± 0.50C.
· Clean
the tissue with tissue paper and care should be taken that there are no air
bubbles in the tube.
· Place
the cell in the sample compartment and measure the reading. After
stabilization, press the set Zero switch, and the Polarimeter should show zero. If not,
rinse the cell with water and repeat the operation until a zero reading is obtained.
· Remove
the cell from the compartment, rinse with 10% w/v sucrose solution, fill it
with the same, clean the cell with tissue paper and place it in the sample
compartment. Ensure that the temperature of the solution is 25 ± 0.50C.
· Measure
the angle of rotation five times and determine the average reading.
· Ensure
that the %RSD of the above five readings is less than 2.0.
· Similarly
measure the angle of rotations for 20%w/v, 30%w/v, 40%w/v and 50%w/v sucrose
solutions.
Accuracy check:
l Check the accuracy of the instrument with
sucrose solution at 589nm and determine through % RSD. The % RSD shall be NMT 2.0%
Calculation:
· Angle
of rotation equivalent to standard concentration (10%, 20%, 30%, 40% & 50%
w/v solutions):
Observes rotation × standard weight
Weight of sucrose
taken in g
Linearity Check:
· Determine the
linearity by using 10%, 20%, 30%, 40% and 50% sucrose solutions and calculate
the correlation coefficient. The correlation
coefficient shall be NLT 0.995.
Acceptance
criteria:
Accuracy with Sucrose (At wavelength 589 nm) |
10%
w/v solution |
+6.665
± 0.1° |
20%
w/v solution |
+13.305
± 0.1° |
|
30%
w/v solution |
+19.930± 0.1° |
|
40%
w/v solution |
+26.530 ± 0.1° |
|
50%
w/v solution |
+33.115 ± 0.1° |
|
Linearity with Sucrose |
Correlation Coefficient Not Less Than 0.995 |
5.3 Maintenance
5.3.1 The sample cell should be cleaned with ethanol/acetone
before performing the test.
5.3.2
Preventive maintenance performed by service engineer
before calibration.
5.3.3
Backup schedule performed every 1 month by IT
personnel.
5.4 Calibration
frequency: 6 months.
6. ANNEXURES
6.1 Annexure I: Autopolarimeter usage logbook
6.2 Annexure II: Calibration Record for Autopolarimeter
REFERENCES
7.1 Operating
instructions manual of Auto polarimeter
7.2 USP
<781> Optical rotation
8.0 REVISION HISTORY
Version no. |
Effective
date |
Summary
of revision |
0.0 |
|
Original Version |
END OF THE DOCUMENT
Logo |
Format
No:
F01-00-QC0XX Page 1 of 1 |
Annexure-I
AUTOPOLARIMETER USAGE LOGBOOK
S.No. |
Date |
Reg. No |
Sample particulars |
Start Time |
End time |
Analyst Sign & Date |
Checked by Sign & Date |
Remarks |
|
|
|
|
|
|
|
|
|
CompanyLogo |
Format
No:
F02-00-QC0XX Page 1 of 2 |
Annexure-II
CALIBRATION RECORD FOR AUTO POLARIMETER
Equipment ID: |
|
Sucrose Make/Grade/Lot No. |
|
Sucrose Expiry date: |
|
||
Balance ID: |
|
Balance Calibration due date: |
|
Thermometer ID: |
|
Thermometer calibration due: |
|
Calibration start date: |
|
Calibration end date: |
|
S.
No. |
Set
Temperature (A) |
Observed value (B) |
Deviation
(A-B) |
1. |
20°C |
|
|
2. |
25°C |
|
|
1) Temperature
Test:
Acceptance criteria: deviation is NMT ±
0.5°C
Result:
Pass / Fail
Performed by Reviewed by
2) Accuracy and Linearity with Sucrose solution
Sno WL(λ) Conc(w/v) A.R.at 25°c Conc(w/v) Conc(w/v) Conc(w/v) Conc(w/v)
1 589 10% 20% 30% 40% 50%
2 589 10% 20% 30% 40% 50%
3 589 10% 20% 30% 40% 50%
4 589 10% 20% 30% 40% 50%
5 589 10% 20% 30% 40% 50%
*Conc.
– Concentration and A.R. Angle of Rotation
Acceptance
criteria: for 10% =+6.665 ±0.1°, for 20%= +13.305 ±0.1°, for 30%= +19.930±0.1°,
for 40%= +26.530 ±0.1°, and for 50%=+33.115 ±0.1°
Logo |
Format
No:
F02-00-QC0XX Page 2 of 2 |
Linearity of Sucrose solution:
S.
No |
Avg. Conc. 10% (w/v) |
Avg. Conc. 20%
(w/v) |
Avg. Conc. 30%
(w/v) |
Avg. Conc. 40%
(w/v) |
Avg. Conc. 50%
(w/v) |
|
|
|
|
|
|
Correlation Coefficient: NLT
0.995 |
|
Result: Pass / Fail
Performed by Reviewed by (Sign& Date)
The results of calibration are “within /out of” acceptance limit and the instrument is ready for further use /requires maintenance.
Next Calibration date:
Remarks:
|
Performed
by |
Reviewed
by HOD/ Designee |
Reviewed
by QA |
Name |
|
|
|
Designation |
|
|
|
Sign & Date |
|
|
|
No comments:
Post a Comment
Don't post spam links