May 31, 2021

SOP for Operation and Calibration of Uv-Vis Spectrophotometer

UV

I           OBJECTIVE 

    To provide a written procedure regarding the Operation and Calibration of UV/VIS (Ultraviolet/Visible) Spectrometer.

II         SCOPE

             This procedure applies to the UV/VIS Spectrometer used in the Quality Control Department.

III        RESPONSIBILITY 

It is the responsibility of Quality control personnel to follow the procedure as given in this SOP.           

IV        DOCUMENT REFERENCE(s)

            SOPs: a) Policy on calibration (Current version of QA 00X)

                        b) Operation and calibration of Analytical Balance (Current version of QC00X)

S.No

Details

Format No.

(Current version)

1

Performance check of UV/VIS spectrophotometer

QC00X/F01

 

        


V         SCHEDULE:

5.1      The performance of the equipment is once every 3 months.

5.2       If the performance check is failed, inform the service engineer.

5.3     In case of any performance check failure, raise a deviation report, and investigate the           Correctness of previous results and the releases.

VI        PROCEDURE  

1.0              Switch “ON” the instrument, computer, and printer.

1.1       Select the UV/Win lab icon, and double-click on the UV/Win lab icon. A display with methods, data region, Result window, graph, start, Auto zero and setup options will appear on the screen.

1.2 Double-click on the methods icon.  It consists of different types of methods, (Scan, TD, WP, Conc. And others) based on their relevant applications.

1.3       Scan: It consists of methods used for wavelength scans between upper and lower ranges. Select the method double click on it and set the start and stop wavelengths along with data interval and number of cycles.

1.4      Wavelength program (WP): It consists of methods used for determining the absorbance. At many wavelengths.  Select the method double-click on it and set the no. of wavelengths along with the no. of cycles.

1.5       Select the method as per requirement and set the method parameters as mentioned above. Open the method and enter the sample name, factor, and sample information.

1.6       Click the “START” button.  Then “Please insert sample Blank” is displayed.

1.7     Then fill the cells with a blank solution and insert them in the sample holder.  After completion of blank correction “Please insert next sample” is displayed.  Then fill the sample cell with the sample solution to be analyzed insert the sample cell into the sample holder and gently close the compartment door.

1.8    At the end of the analysis, a report will be printed automatically. And document the details in the equipment log.

2.0       Calibration of UV/ VIS Spectrometer

2.1       PERFORMANCE CHECK:

2.1.1 Do the baseline and background correction as per the model and operating procedure of the UV/ VIS Spectrometer available in the respective unit.

2.1.2    The following parameters shall be checked for performance verification.

2.1.3    Control of wavelengths.

2.1.4    Control of absorbance.

2.1.5    Limit of stray light.

2.1.6    Wavelength accuracy.

2.1.7    Resolution.

2.2      CONTROL OF WAVELENGTHS:

2.2.1    Preparation of Perchloric acid solution: Dilute 8.5 ml of 70 % perchloric acid to

            100ml with water.

2.2.2    Preparation of Holmium perchlorate solution: Weigh accurately about 0.2 g of

Holmium oxide in a 5 ml volumetric flask, dissolves, and dilutes to volume with a perchloric acid solution. Keep the solution for about 24 hours for solution stabilization.

2.2.3        Scan the Holmium perchlorate solution between 200 nm and 600 nm against perchloric acid as blank.

2.2.4        Check the wavelength absorbance at 241.15 nm, 287.15 nm, 361.5 nm, and 536.3 nm and record observed wavelength absorbance in performance check of UV/VIS Spectrophotometer format.

2.3      CONTROL OF ABSORBANCE:

2.3.1    Preparation of potassium dichromate solution: Dissolve 0.0057 g to 0.0063 g of 

Potassium dichromate was previously dried at 130°C for 4 hours, in 0.005M sulphuric acid and diluted to 100 ml with the same 0.005M sulphuric acid.

2.3.2    Scan the potassium dichromate solution between 200 nm and 400 nm against 0.005M Sulphuric acid as blank.

2.3.3   Check the wavelength absorbance 235 nm, 257 nm, 313 nm, and 350 nm and calculate the specific absorbance.

                                                                 Absorption

                Specific absorbance = ----------------------------------                                                                                                                        Wt. of K2Cr2O7 (in grams)

2.3.4     Record the Specific absorbance in the Performance check of UV/VIS Spectrophotometer format.

2.4       LIMIT OF STRAY LIGHT:

2.4.1    Preparation of potassium chloride solution: Dissolve 1.2 g of potassium chloride,

            Previously dried at 105°C for 4 hours, and dilute to 100 ml with water.

2.4.2    Scan the potassium chloride solution between 190 nm and 220 nm against water as blank.

2.4.3    Check the absorbance between 198 nm and 202 nm, and record the observation in Performance check of UV/VIS Spectrophotometer format.

2.5       WAVELENGTH ACCURACY:

2.5.1    Take the D2 peak spectrum between 660 nm and 650 nm.

2.6       RESOLUTION:

2.6.1    Preparation of Stock solution: Transfer 0.2 ml of Toluene into a 10 ml volumetric

            The flask containing about 5 ml of n-hexane dissolves and dilutes to volume with n-hexane.

2.6.2    Preparation of 0.02 %v/v solution: Transfer 1.0 ml of the above stock solution into a 100 ml volumetric flask containing about 50 ml of n-Hexane, dissolve, and dilute to volume with n-Hexane.

2.6.3    Scan the solution between the range of 250 and 300 nm against n-Hexane as blank. Measure the absorbance of the solution maxima at 269 nm and minima at 266 nm.

                                                 Absorbance at 269 nm

                        Resolution   = -----------------------------

                                                 Absorbance at 266 nm

2.7       ACCEPTANCE CRITERIA:

2.7.1    Control of wavelengths:

                     Wavelength           Tolerance

                        241.15                   ± 1.0
                        287.15                   ± 1.0
                        361.5                     ±1.0
                        536.3                     ± 3.0    

       2.7.2     Control of absorbance:

                      Wavelength          Specific absorbance

                        235 nm                  122.9 to 126.2

                        257 nm                  142.8 to 146.2

                        313 nm                  47.0 to 50.3

                        350 nm                  105.6 to 109.0

 2.7.3       Limit of stray light:

               The absorbance at 200 nm should be more than 2.0

2.7.4       Wavelength accuracy:

                The wavelength accuracy should be between 655.8 and 656.4 nm

2.7.5       Resolution:

               The ratio of the absorbance should be more than 1.5. 

2.8           SCHEDULE:

2.8.1      The performance of the equipment is once in at the time of major spares break down.

2.8.2       If the performance check is failed, inform the service engineer.

2.8.3    In case of any performance check failure, raise a deviation report, and investigate the correctness of previous results and the releases.

2.8.4  Enter the calibration details in the calibration tag and break down details in the Instrument malfunction record.  

REFERENCE:

* Instrument Manual of  Uv-Visible Spectrophotometer

REVISION HISTORY

Version no.

Effective date

Summary of revision

0.0

   DD/MM/YYYY

New SOP


END O F THE DOCUMENT

                                                Prepared By               Checked By                         Approved By

Sign & Date

Name & Department               QA Executive               QC Manager                                QA Manager



                                                                                                                                                                                                                                                                                                                             

 

 

 

 

 

 

No comments:

Post a Comment

Don't post spam links

Calibration of HPLC